Blood Pressure Transducer

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (800) 979-6795

Details

These disposable blood pressure transducers reflect more than 25 years of experience in critical care blood pressure monitoring and it is our goal to continually seek ways to improve accuracy and safety, and to make these blood pressure monitoring products the easiest to use.

Consistent and accurate readings during blood pressure monitoring.

Components that simplify setup.

The transducer can easily be mounted on either the patient, bed or IV pole.

Fully integrated flow-thru component design simplifies filling and debubbling.

Superior fluid path visualization.

3cc or 30cc per hour flow rates.

With a variety of interface cables available, these blood pressure monitoring systems can be connected to most types of monitors.

Specifications

  • Continuous Flow Rate (for models with integral flush device) - 3cc/hr (1cc/hr) or 30cc/hr (10cc/hr) at 300 mmHg
  • Operating Pressure Range - -50 to +300 mmHg
  • Sensitivity - 5 V/V/mmHg, 2% (typically < 1%)
  • Zero drift with time - < 1.0 mmHg/8 hours after 10 min. warm-up to operating temperature
  • Leakage current - <2 A @ 115 Vac rms at 60 Hz
  • Unbalance - 75 mmHg
  • Overpressure protection - -400 to +4000 mmHg
  • Operating Temperature - 15 C to 40C
  • Excitation voltage and frequency - 2 to 10 Vdc; or Vac rms to 5 kHz
  • Operating life - >500 hours
  • Storage temperature - -25C to +65C
  • Defibrillation withstand - 5 discharges/5 minutes of 400 joules @ 50 ohm load
  • Natural frequency - >200 Hz in saline
  • Phase shift - <5 at 5 kHz
  • Output impedance - 270 Ohms to 400 Ohms
  • Input impedance - 270 Ohms to 400 Ohms