DRE Veterinary is now Avante Animal Health

Visit our new website by using the "Proceed to new URL" button.

GE Datex Ohmeda TuffSat Pulse Oximeter

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (800) 979-6795

Details

Designed to deliver accurate SpO2 readings in mobile, fast-paced environments.

The TuffSat is a small, durable, portable pulse oximeter that operates on battery power. It is designed specifically for spot-checking arterial oxygen saturation (SpO2) and pulse rate. This easy-to-use oximeter is ideal for use in any environment.

Features

  • Lightweight, ergonomic design with rubber grip for easy handling.
  • Clear display for quick readings in the field.
  • Relative Perfusion Index (Plr®) helps ensure best probe placement, saving time and increasing accuracy.
  • Battery-operated unit requires four 1.5V alkaline AA batteries.
  • TruTrak data sampling system enables the TuffSat to calculate SpO2 many times each second for reliable readings during times of low perfusion, motion or electrical interference.
  • Low battery indicator.
  • Comes standard with neoprene carrying case with belt clip.

Specifications

Physical

  • Dimensions: 6 in (15 cm) x 2.8 in (7 cm) x 1.2 in (3 cm)
  • Weight: 9 oz (257 g)

Environmental

  • Operating Temperature: -10 to 60° C (14 to 140°F)
  • Operating Relative Humidity: 20 to 95%
  • Operating Atmospheric Pressure: 1060 to 697 hPa
  • Operating Approximate Elevation: -378 to 3048 m (-1240 to 10,000 ft.)
  • Transport Temperature: -40 to 70° C (-40 to 158°F)
  • Transport Relative Humidity: 5 to 95%
  • Transport Atmospheric Pressure: 1060 to 188 hPa
  • Transport Approximate Elevation: -378 to 12.2 km (-1240 to 40,000 ft.)

SpO2

  • Calibration: functional
  • Range: 0 to 100%
  • Accuracy: 70 to 100% ± 2 digits | below 70% unspecified
  • First reading (full accuracy): ≤ 12 seconds
  • Resolution: 1%

Sensors

  • Red LED peak wavelength range: 650 to 670 nm
  • Infrared (IR) LED peak wavelength range: 930 to 950 nm
  • Average power: ≤ 1 mW

Pulse Rate

  • Range: 40 to 255 beats per minute (bpm)
  • Accuracy: 40 to 100: ± 2 bpm | 100 to 255: ±%
  • First reading (full accuracy): ≤12 seconds
  • Resolution: 1 bpm