DRE Veterinary is now Avante Animal Health

Visit our new website by using the "Proceed to new URL" button.

Baxter Colleague

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (800) 979-6795

Details

Features:

  • Fluid delivery applications include parenteral fluids, drugs, electrolytes, total parenteral nutrition (TPN) solutions, whole blood, and blood products
  • Automatically loads administration set tubing
  • Can be used with a variety of source containers
  • Micro and Macro rate range
  • Piggybacking Secondary Medications with Callback Option
  • Does programming modes
  • Configurable air sensor settings
  • user selectable downstream occlusion pressure settings (psig and mmHg)
  • Fluid label library
  • Flow check display indicating downstream in-line resistance
  • Panel lockout with optional Auto Lock function minimizes potential for tampering and IV set removal
  • Configurable Pump Personality feature sets
  • Battery charge level indicator
  • Nurse call
  • RS232 communications port
  • Built-in diagnositcs

Configurable Models:

  • Single Channel
  • Triple Channel

Specifications

  • Dimensions: 10.0" H x 7.75" W x 7.75" D | 254 x 197 x 187 mm
  • Weight: 10.4 lb | 4.7 kg
  • Keep Vein Open (KVO) rate: 0.1-5 ml/hr in 0.1 increments or programmed rate, whichever is less
  • Battery: 12 volt, 2.0 Ah sealed lead acid
  • Battery Life: 5 hrs operating time at 100 mL/hr (with fully charged new battery). 2.5 hrs operating time at 100 mL/hr for the warranted life of the battery
  • Battery Recharge: 8 hrs to recharge to 80% capacity
  • AC Power Requirements: 100-120 VAC, 60 Hz | 220-240 VAC, 50/60 Hz, automatically sensed
  • Power Cord: Approx. 2.9m (9ft)
  • Fuses: 1.6 A Type T (Time Delay), 250 V, two provided
  • Range of programmable:
    • Primary Infusion: 0.1 - 1200 mL
      • 0.1 - 99.9 mL in 0.1 mL/hr increments
      • 1-1200 mL in 1 mL/hr increments
    • Piggyback Infusion: 0.1 - 500 mL/hr
      • 0.1 - 99.9 mL in 0.1 mL/hr increments
      • 1 - 500 mL in 1 mL/hr increments
  • Volume to be Infused: 0.1 - 9999 mL
    • 0.1 - 99.9 mL in 0.1 mL increments
    • 1 - 9999 mL in 1 mL increments
  • Priming Rate: 500 mL/hr
  • Power Consumption: 40 watts maximum
  • Accuracy:
    • 5% or better for any one hr period over the course of 72 hrs at a programmed rate of 100 mL/hr
    • For rates below 1 mL/hr, the accuracy can be maintained at 10% or better for any one hr period of 0.5 mL of delivery, whichever is greater
  • Warranty: 6 months