Masimo Radical 57 Pulse Oximeter

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (800) 979-6795

Details

The Masimo Rad-57 is the world’s most versatile handheld oximeter with a rugged lightweight design ideal for field or hospital settings.

The Rad-57 requires no user calibration and does not need patient cooperation or consciousness. This is a great replacement for existing handheld pulse oximeters. You can get factory-ordered options or field-installed software upgrades to continuous monitoring and spot check measurements of SpHb, SpOC, SpMet and PVI.

Features

  • Helps clinicians reduce the risk of misdiagnosing unsuspected carbon monoxide poisoning
  • Assists clinicians in detecting potentially life-threatening acquired methaemogloninaemia
  • Accurate SpO2 and pulse rate measurements during the most difficult clinical conditions, including during patient motion and conditions of low peripheral perfusion motion
  • Perfusion Index (PI) indicates arterial pulse signal strength and may be used by clinicians to help diagnose during low perfusion
  • User may programme default power-up settings Delivers more than 8 hours of continuous monitoring of SpCO, SpMet, SpO2, Pulse Rate, and Perfusion Index
  • FastSat tracks rapid changes in arterial O2 with unmatched fidelity
  • SmartTone™ beeps in sync with pulse, even under patient motion conditions
  • Up to 72 hours of trending memory of SpCO, SpMet, SpO2, Pulse Rate, and Perfusion Index
  • Sensitivity options of APOD™, Normal and MAX provide flexibilty to support a range of clinical applications

Specifications

Trending

  • Provides 72 hours of trending at 2 second resolution of SpO2, SpCO, SpMet, pulse rate, and perfusion index. Output to PC running Masimo TrendComTM Utility.

SPO2 Modes

  • Averaging mode5 - 2, 4, 8, 10, 12, 14 or 16 seconds
  • Sensitivity - APOD, Normal and Max
  • FastSat 6 - On / Off

Alarms

  • Audible and visual alarms for high and low saturation (SpO2 1% to 100%, SpCO 5% - 40%, SpMet 0.5% to 30%, and pulse rate 25 - 240 bpm) Sensor condition, system failure and low battery alarms

Display / Indicators

  • Data Display - %SpO2, %SpCO, %SpMet, SpCO bar, SpMet bar, pulse rate, perfusion index, low Signal IQ alarm status, alarm silence and battery life
  • Type - LED

Compliance

  • EMC Compliance - EN60601-1-2, Class B
  • Equipment Classification - IEC 60601-1-1
  • Degree of Protection - Type BF-Applied part

Measurement Range

  • SpO2 - 1% - 100%
  • SpCO - 0% - 99%
  • SpMet - 0% - 99.9%
  • Pulse Rate - 25 - 240 (bpm)
  • Perfusion Index - 0.02% - 20%

Oxygen Saturation Accuracy (%SpO2)

  • Saturation - 70% to 100%
  • No Motion
  • Adults, Paediatrics - 2 digits
  • Neonate - 3 digits
  • Motion1
  • Adults, Paediatrics - 3 digits
  • Neonate - 3 digits
  • Low Perfusion2
  • Adults, Paediatrics - 2 digits
  • Neonate - 3 digits

Carboxyhaemoglobin Saturation Accuracy (%SpCO)

  • SpCO - 1% - 40% 3 digits3

Methaemoglobin Saturation Accuracy (%SpMet)

  • SpMet - 1% - 15% 1 digits3

Pulse Rate Accuracy

  • Pulse Rate - 25 - 240 bpm
  • No Motion
  • Adults, Paediatrics, Neonate - 3 digits
  • Motion1
  • Adults, Paediatrics, Neonate - 5 digits
  • Low Perfusion2
  • Adults, Paediatrics, Neonate - 3 digits

Resolution

  • Oxygen Saturation (%SpO2) - 1%
  • Pulse Rate (bpm) - 1 bpm
  • Carboxyhaemoglobin Saturation (%SpCO)
  • Numeric Display - 1%
  • Bar Display - 3, 6, 9, 12, 15, 20, 25, 30, 35, > 40%
  • Methaemoglobin Saturation (%SpMet)
  • Numeric Display - 0.1%
  • Bar Display - 0.5, 1-5, 7.5, 10, 15 > 20%

Accessories