DRE Echo CO2 Capnograph Monitor

Ask About Financing

Available in US only.

Need Equipment Advice?

Ask Our Experts! (800) 979-6795

Details

The DRE Echo CO2 provides cost-effective SPO2 and CO2 monitoring with accuracy and durability.

The Echo CO2 ensures effective capnography monitoring for intubated and non-intubated patients for continuous long term monitoring. It is tailored for mechanically ventilated and non-intubated patients.

Features:

  • 5.7 inch high resolution display for easy reading
  • Lightweight, portable design and user-friendly interface for easy operation
  • Flexible configurations to meet different clinical needs
  • Standard SpO2
  • PR measurement
  • Respironics Loflo sidestream and CAPNOSTAT 5
  • ETCO2 mainstream measurement
  • Real-time parameters measurement display with trend table for easy reviewing
  • Built-in rechargeable Lithium-ion battery for
  • 10 hours continuous working
  • Powerful storage capacity
  • Bi-directional communications with central station by wired or wireless network
  • Nurse call

Specifications

Physical Specifications

  • Size: 200.8mm (L) x 241mm (H) x 189 mm (D)
  • Weight: 3 kg
  • Display: 5.7 inches
  • Color TFT resolution: 640X480
  • Power Supply: 100-240 VAC, 50/60HZ
  • Pmax=70VA FUSE T 1.6AL
  • Battery Type: Lithium-ion
  • Voltage: 14.8 V DC Capacitance: 4,400 mAh
  • Working period: Color TFT 480min
  • Rechargeable period: < 360min
  • Thermal Recorder: (Optional)
  • Record Width: 48 mm
  • Paper Speed: 25mm/s

SpO2

  • Measuring Range: 0 ~ 100 %
  • Alarm Range: 0 ~ 100 %
  • Resolution: 1 %
  • Accuracy
  • Adult (including Pediatric)
  • 2% (70%~100% SpO2)
  • Undefined 0~70% SpO2

Neonate

  • 3% (70%~100% SpO2)
  • Undefined (0~70% SpO2)

Pulse Rate

  • Measuring and Alarm Range: 30 ~ 300bpm
  • Resolution: 1bpm
  • Accuracy: 3bpm
  • Data update period: 2s
  • ISO 9919
  • Respironics CO2

Mainstream

  • CO2 Measurement Range
    • 0 to 150 mm Hg,0 to 19.7%,
    • 0 to 20 kPa (at 760 mm Hg)
  • Rise Time
    • Less than 60 ms Adult/Infant
    • Reusable or Single Patient Use
  • Airway Adapter
  • CO2 Resolution
    • 0.1 mm Hg -- 0 to 69 mm Hg
    • 0.25 mm Hg -- 70 to 150 mm Hg
  • CO2 Accuracy
    • 0-40 mm Hg -- 2 mmHg
    • 41 - 70 mm Hg -- 5% of reading
    • 71 - 100 mm Hg -- 8% of reading
    • 101 - 150 mm Hg -- 10% of reading
    • Above 80 bpm -- 12% of reading
  • CO2 Stability
  • Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
  • Long Term Drift: Accuracy specification will will be maintained over a 120-hour period
  • Respiration Rate Accuracy
    • 1 breath
    • Calibration
  • No routine user calibration required. An airway adapter zero is required when changing to a different style of airway adapter
  • Water Resistance
    • IPX4-Splash-proof (sensor head only)
    • Shock Impact
    • EN60068-2-6 Sinusoidal Vibration
    • EN60068-2-27 Shock
    • EN60068-2-64 Random Vibration
  • Able to withstand repeated 6 foot drops onto tiled floor while operating

Sidestream

  • Sample Rate: 50 m/L per minute
  • CO2 Measurement Range:
    • 0 - 150 mm Hg, 0 to 20 kPa
    • (at 760 mm Hg). Barometric Pressure
    • supplied by Host
  • CO2 Resolution:
    • 0.1 mm Hg 0 to 69 mm Hg
    • 0.25 mm Hg 70 to 150 mm Hg
  • CO2 Accuracy
    • 0 - 40 mm Hg 2 mm Hg
    • 41 - 70 mm Hg 5% of reading
    • 71 - 100 mm Hg 8% of reading
    • 101 - 150 mm Hg 10% of reading
    • Above 80 bpm 12% of reading
  • CO2 Stability
    • Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.
  • Long Term Drift: Accuracy specification will be maintained over a 120-hour period
  • Respiratory Rate Range: 2 to 150 bpm
  • Respiratory Rate Accuracy: 1 breath

Sample Cell/Filter

  • Proprietary single patient use sample cell and inline filter are integrated with the sample line which eliminates contamination of the internal system
  • Nasal Sampling Kits for Non-intubated Patients Adult, pediatric and infant nasal CO2 sampling, nasal CO2 sampling and O2 delivery
  • Adult and pediatric nasal/oral CO2sampling, nasal/oral CO2 sampling and O2 delivery
  • On-Airway Adapter KITS for Intubated Patients
  • Adult/Pediatric with and without dehumidification tubing Pediatric/ Infant, low dead space, with and without dehumidification tubing
  • Taper meets ISO 5356-1
  • Sample Kit Hours of Use
    • Nasal Cannula (all styles) — up to 12 hours
  • On-Airway Adapter Kits without dehumidification tubing — up to 12 hours
  • Sample Cell Detection
  • Insertion automatically turns sampling pump on. Removal automatically turns sampling pump off.
  • Water Resistance IPX4 -Splash-proof(when sample cell is inserted in sample cell receptacle)
  • Shock Impact IEC TR 60721-4-7 Class 7M3(designed to withstand environments subject to significant vibrations or high shock levels)
  • EN60068-2-27 Shock
  • EN60068-2-64 Random vibration