DRE Waveline Touch

  • Product ID: 9460

Details

User-friendly touch-screen patient monitor

  • Measures ECG, SpO2, NIBP, heart rate, temperature and respiration
  • Also available with dual invasive blood pressures, EtCO2 (Sidestream or Mainstream), a battery and a built-in recorder
  • 12.1” color TFT display with a touch-screen interface
  • Multi-lead ECG monitoring
  • ST segment monitoring and 72 hours trend data analysis
  • Built-in battery
  • NIBP list and OxyCRG dynamic view display
  • Drug dose calculations
  • Large font display with 4, 6 and 8 waveforms shown
  • Built-in thermal recorder
  • Networkable with a central monitoring system via a wireless or wired network
  • Compact design convenient for mobile monitoring
  • Multi-lingual settings
  • Includes veterinary accessories for parameters ordered

Each monitor will automatically include the CO2 and IBP ports. These parameters are listed as options and are NOT standard. Your veterinary or research facility has the ability to upgrade to any of these parameters at a later time without having to send the monitor into DRE! Call a DRE Account Manager for further details.

Specifications

Physical Dimensions & Weight
Main Unit -
  • 21.9” H x 13.2” W x 9.23” D (555 mm H x 335 mm W x 235 mm D)
  • Weight - 11bs (5kg)
  • Application
  • Veterinary patients
  • Operation Environment
  • Power Source - External AC power or internal battery
  • AC Power - 100-240VAC, 50/60Hz, <150VA
  • Battery - Rechargeable Lead-Acid; Operating time under the normal condition (1 hour); Operating time after the first alarm of low battery (10 min.)
Temperature
  • Working - 5-40 C
  • Storage - -20-65 C
  • Relative Humidity
  • Working - 30-75%
  • Storage - ≤80%
  • Altitude
  • Operating Altitude - Up to 5,000 meters
  • Hyperbaric Pressure - Up to 405.3 kPa
Performance Specifications
  • Display - 12.1” color TFT, Touch Screen
  • Resolution - 800 x 600 pixels
  • Trace - 4, 6 or 8 waveforms ECG (I, II, III, aVR, aVL, aVF, V1-V6), PLETH, RESP, IBPx2, ETCO2
  • Indicator - Alarm indicator; Power indicator QRS beep and alarm sound
  • Trend Time - 72 hours
  • Recorder - Built-in, thermal array, 2 channels
  • Record - Width 48mm
  • Recorder - Paper 50mm
  • Record - Speed 25mm/s, 50mm/s
  • OxyCRG - Combines HR, SpO2, and resp. trends into a single graph ECG
  • Input - 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice - I, II, III, aVR, aVF, aVL, V
  • Gain Choice - x0.125, x0.25, x0.5, x1.0, x2.0
Filter:
  • Diagnostic mode - 0.05 – 100Hz
  • Monitoring mode - 0.05 – 75Hz
  • Surgical mode - 1 – 20Hz
  • Frequency Characteristic - 0.05-40 Hz
  • ECG Waveforms - 7 channels
  • Penetration Voltage - 4000VAC 50/60Hz
  • Sweep Speed - 12.5, 25 and 50 mm/s
  • HR Display Range - 30-300bpm
  • Accuracy - 1% or 1bpm, whichever is greater
  • S-7 Segment Detection - 7 channels
  • Measurement Range - -2.0mv – +2.0mv
  • Arrhythmia Analysis - 13 types
  • Alarm Limit Range Setting - Upper limit 80-400bpm; Lower limit 20-150bpm
  • RESP
  • Measure Method - RA-LL impedance
  • Range - 0-120 rpm
  • Accuracy - + 3 rpm
  • Alarm Upper-lower Limit Setting - Upper limit 6-120 rpm, Lower limit 3-120 rpm
  • Sweep Speed - 12.5 and 25 mm/s
  • NIBP
  • Measuring Technology - Automatic oscillating measurement
  • Cuff Inflating - <30s (0 – 300mmHg, standard adult cuff
  • Measuring Period - AVE<40s
  • Mode - Manual, Auto, STAT
  • Measuring Interval in AUTO Mode - 2 min-4 hrs
  • Pulse Rate Range - 30-250 (bpm)
  • Measuring Range - Adult/Pediatric Mode, SYS 40-250 (mmHg), DIA 15-200 (mmHg); Neonatal Mode, SYS 40-135 (mmHg), DIA 15-100 (mmHg)
  • Resolution - 1mmHg
  • Accuracy - Pressure, Maximum Mean error 5mmHg, Max. Standard deviation 8mmHg
  • Overpressure Protection - Adult Mode 280(mmHg), Neonatal Mode 150 (mmHg)
  • Alarm Limit Setting - SYS 50-240 mmHg, DIA 15-180 mmHg
Temp:
  • Range - 0 – 50šC (32 – 122šF)
  • Accuracy - Without sensor
  • Display Resolution - 0.1 C
  • Alarm Upper-lower Limit Setting - Upper limit 0-50 C, Lower limit 0-50 C
  • Channel - 2 channels, provide T1, T2, ΔT
  • Alarm Limit Setting - 10-50 C
  • SPO2 (Nellcor)
  • ASpO2 Anti-motion SpO2
  • SpO2% Range 0-100%
  • SpO2 Accuracy + 2% 70-100%, nonmotion; + 3% 70-100%, motion
  • Pulse Rate Range 30-250 bpm
  • Pulse Rate Accuracy 2 bpm (non-motion); 3 bpm (motion)
  • Alarm Upper-lower Limit Setting: Upper limit 70-100%, Lower limit 70-100%
  • SpO2 Probe - Red Light LED Wavelength 660nm 5nm, Infrared Light LED
  • Wavelength - 940nm 10nm
  • IBP (Option)
  • Measurement Range - -10-300mmHg
  • Channel - 2 channels
  • Pressure Transducer Sensitivity, 5μV/V/mmHg
  • Impedance Range - 300-3000Ω
  • Transducer Sites - ART, PA, CVP, RAP, LAP, ICP
  • Resolution - 1mmHg
  • Accuracy - 1mmHg or 2%, whichever is greater
  • Alarm range - -10-300mmHg
  • EtCO2 (Option) (Respironics)
  • Mode of Sampling - Sidestream
  • Principle of Operation -Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
  • CO2 Measurement Range - 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
  • CO2 Calculation Method - BTPS (Body Temperature Pressure Saturated)
  • CO2 Resolution - 0.1mmHg(0-69mmHg); 0.25mmHg(70-150mmHg)
  • CO2 Accuracy - 0-40 mmHg 2 mmHg; 41-70 mmHg 5% of reading; 71-100 mmHg 8% of reading; 101-150 mmHg 10% of reading; Above 80 breath per minute 12% of reading
  • Sampling Rate - 100Hz
  • Respiration Rate - 2-150 bpm
  • Respiration Rate Accuracy - 1 breath
  • Response Time - <3 seconds - includes transport time and rise time
  • Inspired CO2 Measurement Range - 3-50 mmHg