DRE Waveline Pro

  • Product ID: 2012

Details

Superior flexibility and performance with a premium flat panel display is found in DRE's newest version of the Waveline monitor series. The DRE Waveline ProVet can be used on small, medium, or large animals. An applicable monitor for veterinary surgical suites and research facilities.

The DRE Waveline Pro displays as many as eight colorful waveforms on a crisp 15” high resolution display. It is lightweight, portable and measures multiple parameters, including ECG (3 or 5 lead) with ST segment monitoring, resp., dual temp., SpO2 (Nellcor) and NIBP. Configurations are available that measure capnography, dual channel IBP, and anesthetic agents (five agents, N2O, EtCO2 (Oridion)), and that have a built-in thermal printer.

Multiple configurations available:

  • Standard Parameters: ECG, Pulse Oximetry (Nellcor), 2 x Temperature, NIBP, Respiration, and Printer
  • Optional Parameters: Invasive Blood Pressures, and Anesthetic Agent Module
  • Comes with standard veterinary accessories

Features:

  • EtCO2 (mainstream or sidestream) (Respironics)
  • Multi-lead simultaneous ECG monitoring
  • Automatically set alarm limits
  • Graphical and tabular trending
  • Color-coded alarms
  • Patient data entry
  • Battery backup (two removable batteries)
  • Built-in thermal recorder
  • Networking capability
  • Large (15”), high resolution flat panel display
  • Available with a built-in thermal printer

Specifications

Physical Dimensions & Weight:
  • Base Unit: 12.1”/309 mm (H) x 14.4”/365 mm (W) x 6.3”/159 mm (D)
  • Weight: 17.6 lbs/8.0 kgs (including anesthetic agents module and batteries)
Application:
  • Small, Medium, and Large animal
Performance Specifications:
  • Display: 15” color TFT, Resolution 1024 x 768
  • Trace: 8 waveforms
  • Indicators:
  • Alarm indicator
  • Power indicator
  • QRS beep and alarm sound
  • Trend time: 1 – 48 hours
  • Recorder:
  • Built-in, thermal array, 2 channels
  • Record width: 1.9”/48 mm
  • Recorder paper: 2”/50 mm
  • Record speed: 25 mm/s, 50 mm/s
ECG:
  • Input: 5-lead ECG cable and standard AAMI line for connection
  • Lead Choice: I, II, III, aVR, aVF, aVL, V, TEST
  • Gain Choice : x0.5, x1, x2, x4
  • Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
  • ECG Waveforms: 7 channels
  • Penetration Voltage: 4000VAC 50/60Hz
  • Sweep Speed: 12.5, 25 and 50 mm/s
  • HR Display Range: 30 – 300bpm
  • Accuracy: 1 bpm or 1%, whichever is greater
  • Alarm Limit Range Setting: Upper limit 100 – 200 bpm; Lower limit 30 – 100 bpm
RESP:
  • Measure Method: RA-LL impedance
  • Range: 0 – 120 rpm
  • Accuracy: 3 rpm
  • Alarm Limit Setting: Upper limit 6 – 120 rpm: Lower limit 3 – 120 rpm
  • Sweep Speed: 12.5 and 25 mm/s
TEMP:
  • Range: 25 – 50 (C)
  • Accuracy:
  • 0.2 C (25.0 – 34.9 C)
  • 0.1 C (35.0 – 39.9 C)
  • 0.2 C (40.0 – 44.9 C)
  • 0.3 C (45.0 – 50.0 C)
  • Display Resolution: 0.1 C
  • Alarm Limit Setting: Upper limit 0 – 50 C, lower limit 0 – 50 C
  • Channel: 2 channels
  • SpO2: (Nellcor)
  • ASpO2: Anti-motion SpO2
  • SpO2% Range: 0 – 100%
  • SpO2 Accuracy: 2% (70 – 100%, non-motion) 3% (70 – 100%, motion)
  • Pulse Rate Range: 30 – 250 bpm
  • Pulse Rate Accuracy: 2 bpm (non-motion) 3 bpm (motion)
  • Alarm Limit Setting: Upper limit 70 – 100%, lower limit 70 – 100%
  • SpO2 Probe Y-site:
  • Red light LED wavelength – 660nm 5nm
  • Infrared light LED wavelength – 940nm 10nm
NIBP:
  • Measuring Technology: Automatic oscillating measurement
  • Cuff Inflating: <30s (0 – 300 mmHg, standard adult cuff)
  • Measuring Period: AVE<40s
  • Mode: Manual, Auto
  • Measuring Interval in AUTO Mode: 2 min – 4 hrs
  • Pulse Rate Range: 30 – 250 (bpm)
  • Measuring Range:
  • Adult/Pediatric Mode
    • SYS: 40 – 250 (mmHg) DIA :15 – 200 (mmHg)
  • Neonatal Mode
    • SYS: 40 – 135 (mmHg) DIA : 15 – 100 (mmHg)
  • Resolution: 1mmHg
Accuracy:
  • Maximum Mean error: 5mmHg
  • Maximum Standard deviation: 8mmHg
  • Overpressure Protection:
  • Adult Mode: 300(mmHg)
  • Neonatal Mode: 160 (mmHg)
  • Alarm Limit Setting:
  • - SYS: 50 – 240 mmHg
  • - DIA: 15 – 180 mmHg
IBP:
  • Measurement Range: -50 – 300mmHg
  • Channel: 2 channels
  • Pressure Transducer: Sensitivity, 5μV/V/mmHg
  • Impedance Range: 300 – 3000Ω
  • Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
  • Unit: mmHg/kPa selectable
  • Resolution: 1 mmHg
  • Accuracy: 1 mmHg or 2%, whichever is greater
  • Alarm Range: -10 – 300 mmHg
EtCO2 (Respironics) (mainstream or sidestream):
  • CO2 Measurement Range: 0 – 99 mmHg
Accuracy:
  • 2 mmHg (0 – 38 mmHg)
  • 39-99 mmHg 5% of reading + 0.08% for every 1 mmHg (above 38 mmHg)
  • Sampling Rate: 50 ml/min
  • Initialization Time: 30 seconds (typical), reaches 5% steady-state accuracy within 3 minutes
  • Respiration Rate: 0 – 150 breaths/min
  • Mode: Adult, neonate
Anesthetic Agents:
  • Industry standard 802.11b/g wireless network
Power:
  • Source: External AC power or internal battery
  • AC Power: 100 – 240VAC, 50/60Hz, 150VA
  • Battery: Built-in and rechargeable
  • Charge Time: 4 hours
Environmental Specifications:
Temperature:
  • Operating: 5 – 40 C
  • Storage: -20 – 65 C
  • Humidity range:
  • Operating: ≤80 %
  • Storage: ≤80 %