Baxter Colleague
- Product ID: 617
- Condition: DRE Refurbished
Details
Features:
- Fluid delivery applications include parenteral fluids, drugs, electrolytes, total parenteral nutrition (TPN) solutions, whole blood, and blood products
- Automatically loads administration set tubing
- Can be used with a variety of source containers
- Micro and Macro rate range
- Piggybacking Secondary Medications with Callback Option
- Does programming modes
- Configurable air sensor settings
- user selectable downstream occlusion pressure settings (psig and mmHg)
- Fluid label library
- Flow check display indicating downstream in-line resistance
- Panel lockout with optional Auto Lock function minimizes potential for tampering and IV set removal
- Configurable Pump Personality feature sets
- Battery charge level indicator
- Nurse call
- RS232 communications port
- Built-in diagnositcs
Configurable Models:
- Single Channel
- Triple Channel
Specifications
- Dimensions: 10.0" H x 7.75" W x 7.75" D | 254 x 197 x 187 mm
- Weight: 10.4 lb | 4.7 kg
- Keep Vein Open (KVO) rate: 0.1-5 ml/hr in 0.1 increments or programmed rate, whichever is less
- Battery: 12 volt, 2.0 Ah sealed lead acid
- Battery Life: 5 hrs operating time at 100 mL/hr (with fully charged new battery). 2.5 hrs operating time at 100 mL/hr for the warranted life of the battery
- Battery Recharge: 8 hrs to recharge to 80% capacity
- AC Power Requirements: 100-120 VAC, 60 Hz | 220-240 VAC, 50/60 Hz, automatically sensed
- Power Cord: Approx. 2.9m (9ft)
- Fuses: 1.6 A Type T (Time Delay), 250 V, two provided
- Range of programmable:
- Primary Infusion: 0.1 - 1200 mL
- 0.1 - 99.9 mL in 0.1 mL/hr increments
- 1-1200 mL in 1 mL/hr increments
- Piggyback Infusion: 0.1 - 500 mL/hr
- 0.1 - 99.9 mL in 0.1 mL/hr increments
- 1 - 500 mL in 1 mL/hr increments
- Primary Infusion: 0.1 - 1200 mL
- Volume to be Infused: 0.1 - 9999 mL
- 0.1 - 99.9 mL in 0.1 mL increments
- 1 - 9999 mL in 1 mL increments
- Priming Rate: 500 mL/hr
- Power Consumption: 40 watts maximum
- Accuracy:
- 5% or better for any one hr period over the course of 72 hrs at a programmed rate of 100 mL/hr
- For rates below 1 mL/hr, the accuracy can be maintained at 10% or better for any one hr period of 0.5 mL of delivery, whichever is greater
- Warranty: 6 months