The DRE Echo CO2 provides cost-effective SPO2 and CO2 monitoring with accuracy and durability. The Echo CO2 ensure effective capnography monitoring for intubated and non-intubated patients for continuos long term monitoring. It is tailored for mechanically ventilated and non-intubated ... [Full Description]
The DRE Echo CO2 provides cost-effective SPO2 and CO2 monitoring with accuracy and durability. The Echo CO2 ensure effective capnography monitoring for intubated and non-intubated patients for continuos long term monitoring. It is tailored for mechanically ventilated and non-intubated patients.FEATURES
for easy operation
ETCO2 mainstream measurement
with trend table for easy reviewing
10 hours continuous working
wired or wireless network
Technical Specifications Mainstream Sidestream Specifications subject to change without notice.
SpO2
± 2% (70%~100% SpO2)
Undefined 0~70% SpO2
± 3% (70%~100% SpO2)
Undefined (0~70% SpO2)
Measuring and Alarm Range: 30 ~ 300bpm
Resolution: 1bpm
Accuracy: ± 3bpm
Data update period: 2s
ISO 9919Respironics CO2
0 to 150 mm Hg,0 to 19.7%,
0 to 20 kPa (at 760 mm Hg)
Less than 60 ms Adult/Infant
Reusable or Single Patient Use
Airway Adapter
0.1 mm Hg -- 0 to 69 mm Hg
0.25 mm Hg -- 70 to 150 mm Hg
0-40 mm Hg -- ±2 mmHg
41 - 70 mm Hg -- ± 5% of reading
71 - 100 mm Hg -- ± 8% of reading
101 - 150 mm Hg -- ±10% of reading
Above 80 bpm -- ±12% of reading
Short Term Drift: Drift over four hours shall
not exceed 0.8 mm Hg max.
Long Term Drift: Accuracy specification will
will be maintained over a 120-hour period
±1 breath
No routine user calibration required. An
airway adapter zero is required when
changing to a different style of airway adapter
IPX4-Splash-proof (sensor head only)
EN60068-2-6 Sinusoidal Vibration
EN60068-2-27 Shock
EN60068-2-64 Random Vibration
Able to withstand repeated 6 foot drops onto tiled floor while operating
0 - 150 mm Hg, 0 to 20 kPa
(at 760 mm Hg). Barometric Pressure
supplied by Host
0.1 mm Hg 0 to 69 mm Hg
0.25 mm Hg 70 to 150 mm Hg
0 - 40 mm Hg ±2 mm Hg
41 - 70 mm Hg ±5% of reading
71 - 100 mm Hg ±8% of reading
101 - 150 mm Hg ±10% of reading
Above 80 bpm ±12% of reading
Short Term Drift: Drift over four hours shall
not exceed 0.8 mm Hg max.
Long Term Drift: Accuracy specification will be
maintained over a 120-hour period
Proprietary single patient use sample cell and
inline filter are integrated with the sample
line which eliminates contamination of the
internal system
Adult, pediatric and infant nasal CO2
sampling, nasal CO2 sampling and O2 delivery
Adult and pediatric nasal/oral CO2sampling,
nasal/oral CO2 sampling and O2 delivery
Adult/Pediatric with and without
dehumidification tubing Pediatric/
Infant, low dead space, with and without
dehumidification tubing
Taper meets ISO 5356-1
Nasal Cannula (all styles) — up to 12 hours
On-Airway Adapter Kits without
dehumidification tubing — up to 12 hours
Insertion automatically turns sampling pump
on. Removal automatically turns sampling
pump off.
(when sample cell is inserted in sample
cell receptacle)
(designed to withstand environments subject to significant
vibrations or high shock levels)
EN60068-2-27 Shock
EN60068-2-64 Random vibration